The self-constructed new STXM endstation has successfully realized high-resolution (30nm) imaging, and the scanning imaging efficiency has been improved by ten times compared with the old endstation through the bidirectional scanning technology and FPGA timing control scheme. [J. Synchrotron Rad. 2021, 28, 512]
New STXM chamber appearance, internal structure and control interface
New STXM star pattern imaging (30nm innermost stripe) and large field of view imaging of cancerous tissue (86um x 76um, 25nm step)