1.Vector-magnetic field endstation
• Magnetic field: 0.9T (maximum, spindle directions); 0.6T(maximum, arbitrary directions)
• Temperature working range: 15-400K
• Sample stage movement freedom: vertical direction and rotate.
• TEY and TFY
• In-situ bias test kit
2.Superconductive magnetic field endstation
• Magnetic field: 9T (maximum, horizontal plane along X-ray); 4T(maximum, vertical to X-ray at horizontal plane);2.5T(maximum, arbitrary directions in horizontal plane)
• Temperature working range: 4-325K
• Sample stage movement freedom: vertical direction and rotate.
• TEY
• Focus spot size:80*60um
3.Nano-ARPES endstation: Angle-resolved photoelectron spectroscopy with spatial resolution capability
• Sample testing chamber rotation range:±15°
• Sample rotation range:±15°
• Sample spot size:200nm×200nm
• Spectrometer energy resolution:≤ 10meV@91eV
4.Spin-AREPS endstation: Probing the electronic spin structure of materials
• Spectrometer energy resolution:≤ 30meV@91eV
• Momentum resolution: 0.5°;
• Total spin analyzer efficiency: > 40
5.Picosecond time resolution testing chamber
• Synchronize the accelerator frequency
• In-situ magnetic field, microwave filed
• Minimum time delay adjustment: 5ps
• TEY and TFY
• Temperature working range: 10-400K
附件下载: